Specifications
R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 78
CALCulate<1|2>:MARKer<1 to 4>:Y <numeric_value>
This command positions the selected marker to the indicated Y position in the
selected measurement window. The units for this command are dBc/Hz.
Example
CALC:MARK:Y?
'Outputs the measured value of marker 1 in screen A.
Characteristics
*RST value: -
SCPI: device-specific
3.3.6 CALCulate:SNOise Subsystem
The CALCulate:SNOise subsystem allows spot noise measurement points to be set,
and results returned.
Note that for all spot noise commands the suffix for the CALCulate must be 1.
CALCulate<1|2>:SNOise<1 to 4>:AOFF
CALCulate<1|2>:SNOise<1 to 4>:STATe <Boolean>
CALCulate<1|2>:SNOise<1 to 4>:X <numeric_value>
CALCulate<1|2>:SNOise<1 to 4>:Y <numeric_value>
CALCulate<1|2>:SNOise<1 to 4>:AOFF
This command switches off all active spot noise markers in the specified
measurement window.
Example:
CALC1:SNO:AOFF
’Switches off all spot noise markers in the screen A window.
Characteristic
*RST value: -
SCPI: device-specific
CALCulate<1|2>:SNOise<1 to 4>:STATe <Boolean>
This command switches on or off the currently selected spot noise marker in the selected
measurement window. If no indication is made, marker 1 is selected automatically.
Example
CALC1:SNO1:STATE ON
'Switches the screen A marker ON.
Characteristics
*RST value: 1
SCPI: device-specific