Specifications

R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 69
Each limit line can be assigned a name. An explanatory comment can also be given for
Example.
Definition and use of a new limit line 5 for trace in the Noise Figure trace screen with
the following features:
F upper limit line
F
5 ref. values: 126 MHz/-40 dB, 127 MHz/-40 dB, 128 MHz/-20 dB, 129 MHz/-40 dB,
130 MHz/-40 dB
Definition of the line:
Defining the name: CALC:LIM5:NAME 'TEST1'
Entering the comment: CALC:LIM5:COMM 'Upper limit line'
Associated trace in screen A: CALC:LIM5:TRAC 2
Defining the X-axis values: CALC:LIM5:CONT 126MHZ, 127MHZ, 128MHZ,
129MHZ, 130MHZ
Defining the y values: CALC:LIM5:UPP -40, -40, -30, -40, -40
The definition of the safety margin and shifting in X- and/or Y-direction can take place
as from here (see commands below).
Switching on and evaluating the line
Switching on the line CALC:LIM5:UPP:STAT ON
Switching on the limit CALC:LIM5:STAT ON
Starting a new measurement with synchronization: INIT;*WAI
Querying the limit check result: CALC:LIM5:FAIL?
CALCulate:LIMit<1 to 8>:CLEar: [:IMMediate] <Boolean>
CALCulate:LIMit<1 to 8>:COMMent <string>
CALCulate:LIMit<1 to 8>:COPY 1 to 8 | <name>
CALCulate:LIMit<1 to 8>:DELete
CALCulate:LIMit<1 to 8>:FAIL?
CALCulate:LIMit<1 to 8>:NAME <name>
CALCulate:LIMit<1 to 8>:STATe ON | OFF
CALCulate:LIMit<1 to 8>:TRACe < numeric_value >
CALCulate:LIMit<1 to 8>:CLEar: [:IMMediate] <Boolean>
This command clears the STATus:QUEStionable:LIMit event register for the
selected limit line.
Example
CALC:LIM:CLE
'Clears the event register.
Characteristics
*RST value: -
SCPI: conforming
This command is an event and is therefore not assigned an *RST value.