Specifications

R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 67
Example
CALC1:DELT:TRAC 2
'
Assigns marker 1 in screen A to trace 2.
CALC2:MARK:TRAC 5
'Assigns marker 1 in screen B to trace 5.
Characteristics
*RST value: 1
SCPI: device-specific
CALCulate<1|2>:DELTamarker<1 to 4>:X <numeric_value>
This command positions the selected delta marker to the indicated frequency or
time in the selected measurement window.
Example
CALC1:DELT:X 2MHZ"
Positions marker 1 in screen A to time 2 MHz.
Characteristics
*RST value: -
SCPI: device-specific
CALCulate<1|2>:DELTamarker<1 to 4>:Y <numeric_value>
This command positions the selected delta marker to the indicated Y position in the
selected measurement window. The units for this command are dBc/Hz.
Example
CALC1:DELT:Y?
'Outputs the measured value of marker 1 in screen A.
Characteristics
*RST value: -
SCPI: device-specific
3.3.2 CALCulate:EVALuation Subsystem
CALCulate<1|2>:EVALuation
CALCulate<1|2>:EVALuation:STARt
CALCulate<1|2>:EVALuation:STOP
CALCulate<1|2>:EVALuation
This command specifies whether Residual noise values are calculated over the
entire trace or within a specified frequency range. When switched ON the residual
noise values are calculated over the range specified by
CALCulate<1|2>:EVALuation:STARt and CALCulate<1|2>:EVALuation:STOP.
When switched OFF the results are calculated over the entire trace.