S-K40 Measurements & Settings
Software Manual 1161.8209.42 - 04 52
2.4.3 Carrier Frequency Offset Table
From: This column lists the start frequency of each sub-band.
To: This column lists the stop frequency of each sub-band.
RBW: This column lists the resolution filter bandwidth used for each sub-band. Enter
values in steps of 1/3/10.
Range: 0.1 % .. 30 % of the start frequency in that row.
Average: This column lists the number of sweeps to average over for each sub-
band.
Range: 1 .. 10000
FFT: It allows the user to choose whether to use the FFT Resolution Filter or the
conventional filter for each decade.
FFT is only available for RBW values between 1 Hz and 30 kHz.
Meas
Time:
Shows the estimated measurement time for each sub-band. Note this time is for
the measurement only and does not include processing time.
An overall measurement time is displayed at the bottom of the table showing the
overall measurement time for the selected sub-bands.
2.4.4 Preset Settings
The Preset Settings group provides a short-cut method quickly configuring the settings
in the Carrier Frequency Offset table.
Settings any of the settings in the Preset Settings group sets all the appropriate
settings in the carrier offset table.
Note that when the Sweep Mode parameter is set to any value other than Manual then
the Preset Settings group is disabled