Specifications

R
&S
F
S-K40 Measurements & Settings
Software Manual 1161.8209.42 - 04 47
Remote DISP:WIND1:TRAC1:SMO:APER
2.3.5.3 Smoothing Type
Smoothing Type whether linear or logarithmic smoothing is to be used when trace
smoothing is performed.
For more details regarding smoothing and the algorithm used see section Trace Menu
Remote DISP:WIND1:TRAC1:SMO:TYPE 1
2.3.6 Residual Calculations
2.3.6.1 Use Meas Settings
Use Meas Settings specifies whether to use the whole measurement range or the user
defined evaluation range for the residual calculations.
If the Use Meas Settings is switched on then the Eval From and To become disabled and
residual calculations are performed across the complete range of the measurement results
If the Use Meas Settings is switched off then the Eval From and To become enabled
and allow the range over which residual calculations are performed to be specified.
Remote CALC1:EVAL:STAT
2.3.6.2 Eval From
Eval From specifies the start of the measurement range for which residual calculations
are to be performed
The minimum value that can be specified for the Eval From setting is the value of the X
Axis Start setting
The maximum value that can be specified for the Eval From setting is the value of the
X Axis Stop setting
When the X Axis Start or X Axis Stop settings are modified the Eval From will be
automatically adjusted to ensure that it is not outside the measurement range