Specifications

R
&S
F
S-K40 Measurements & Settings
Software Manual 1161.8209.42 - 04 45
2.3.4 Display Settings
2.3.4.1 X Axis Start
X Axis Start parameter specifies the min frequency for the X axis.
When X Axis Start changes the Start parameter in Measurement Settings view gets
updated accordingly.
Remote SENS:FREQ:STAR
2.3.4.2 X Axis Stop
X Axis Stop parameter specifies the max frequency for the X axis.
When X Axis Stop changes the Stop parameter in Measurement Settings view gets
updated accordingly.
Remote SENS:FREQ:STOP
2.3.4.3 Autoscale Once
If the Autoscale Once is ON, the y-axis scaling is calculated from results.
The autoscaling will only be done once in the first sweep. The subsequence sweeps
will not auto scale the y-axis.
When Autoscale Once is ON, Y Axis Top and Range parameters will be greyed out.
When it is OFF, Y Axis Top and Range parameters will be editable.
Remote DISP:WIND1:TRAC1:Y:SCAL:AUTO