Specifications
R
&S
F
S-K40 Measurements & Settings
Software Manual 1161.8209.42 - 04 44
2.3.2.6 Level Tolerance
Level Tolerance parameter is an offset relative to the Level. It is used to verify the
Level of the input signal.
Level Tolerance specifies the maximum and mimum deviation from the specified Level
setting that the input signal may vary by and still pass the verification i.e. measured
level between (Level+Level_Tolerance) and (Level-Level_Tolerance) is accepted.
Remote SENS:RLEV:VER:TOL
2.3.3 AFC Settings
2.3.3.1 Track Frequency
Track Frequency parameter enables or disables the Signal Frequency tracking
mechanism during the measurement.
This parameter is only available when the Verify Frequency & Level parameter is
enabled
Remote SENS:FREQ:TRAC
2.3.3.2 Track Level
Track Level parameter enables or disables the Signal Level tracking mechanism during
the measurement.
This parameter is only available when the Verify Frequency & Level parameter is
enabled.
Remote SENS:POW:TRAC