Specifications

R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 102
3.15.1.2 STATus:QUEStionable Register
This register comprises information about indefinite states which may occur if the unit
is operated without meeting the specifications. It can be queried by commands
STATus:QUEStionable:
CONDition?
and
STATus:QUEStionable[:EVENt]
?.
Bit No Meaning
0 to 2 These bits are not used
3 POWer
This bit is set if a questionable power occurs
4 TEMPerature
This bit is set if a questionable temperature occurs.
5 FREQuency
The bit is set if a frequency is questionable
6 to 7 These bits are not used
8 CALibration
The bit is set if a measurement is performed uncalibrated (= ^ label "UNCAL")
9 LIMit (device-specific)
This bit is set if a limit value is violated (see also section STATus:QUEStionable:LIMit
Register). Note: Limit register is associated with limit lines for the Spectrum Mask
measurement only.
10 LMARgin (device-specific)
This bits are not used within R&S FS-K40
11 SYNC (device-dependent)
This bits are not used within R&S FS-K40
12 ACPLimit
This bits are not used within R&S FS-K40
13 to 14 These bits are not used
15 This bit is always 0
3.15.1.3 STATus:QUEStionable:LIMit Register
This register comprises information about the observance of limit lines in the
corresponding
measurement window (LIMit 1 corresponds to Screen A, LIMit 2 to Screen B). It can be
queried with
commands
STATus:QUEStionable:LIMit<1|2>:CONDition?
and
STATus:QUEStionable:
LIMit<1|2>[:EVENt]?
.
Note that no limit lines are displayed in screen B and as such all bits in the LIMit2
register will always be set to 0.