Specifications
R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 92
[SENSe]:POWer:RLEVel:VERify:TOLerance <numeric_value>
The remote control command is used to specify the power tolerance for the verification of
the signal. If the signal level varies from the specified level by a value greater than the
tolerance then the verification will fail.
The setting for this parameter only has an effect if the
[SENSe]:FREQuency:VERify:STATe or [SENSe]:POWer:RLEVel:VERify:STATe
command is set to ON.
Example
SENS:POW:RLEV:TOLerance 5DB
'Sets the level tolerance for the verification measurement to
1 dB
Characteristics
*RST value: 10 dB
SCPI: conforming
[SENSe]: POWer:TRACk <Boolean>
The remote control command switches the automatic level control on and off.
Example
SENS:POW:TRACk ON
'Sets the automatic level control to ON.
Characteristics
*RST value: ON
SCPI: conforming
[SENSe]:SWEep:COUNt <numeric_value>
The remote control command is used to specify the number of sweeps.
Example
SENS:SWEep:COUNt 64
'Sets the number of sweeps to 64.
Characteristics
*RST value: 1
SCPI: conforming
[SENSe]:SWEep:FORWard <Boolean>
The remote control command specifies the sweep direction. When switched on the
sweep direction is from the start frequency to the stop frequency. When switched
off the sweep direction is reversed