Specifications
R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 90
[SENSe]:FREQuency:TRACk <Boolean>
The remote control command switches the automatic frequency control on and off.
Example
SENS:FREQ:TRACk ON
'Sets the automatic frequency control to ON.
Characteristics
*RST value: ON
SCPI: conforming
[SENSe]:FREQuency:VERify[:STATe] <Boolean>
The remote control command is used to specify whether frequency and level
verification are to be performed before a phase noise measurement.
This command is synonymous with [SENSe]:POWer:RLEVel:VERify[:STATe]
Example
SENS:FREQ:VER 1
'Specifies that frequency and level verification is to be
performed
Characteristics
*RST value: 1
SCPI: conforming
[SENSe]:FREQuency:VERify:TOLerance: <numeric_value>
The remote control command is used to specify the absolute frequency tolerance
for the verification of the signal. If the signal frequency varies from the specified
centre frequency by a value greater than the tolerance then the verification will fail.
The setting for this parameter only has an effect if the
[SENSe]:FREQuency:VERify:STATe or [SENSe]:POWer:RLEVel:VERify:STATe
command is set to ON.
The tolerance used during the measurement is the highest of the Absolute & the
relative tolerance values.
Example
SENS:FREQ:VERify:TOL 1KHZ
'Sets the absolute frequency tolerance for the verification
measurement to 1 kHz
Characteristics
*RST value: 1 PCT
SCPI: conforming