Specifications

R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 88
3.12 SENSe Subsystem
The SENSe command is used to set and get the values of parameters in the remote
instrument. The get variant of the SENSe command differs from set in that it takes no
parameter values (unless otherwise stated) but is followed by the character ‘?’ and will
return the parameter’s value in the same format as it is set.
e.g
SENS:FREQ 10GHZ - sets the frequency to 10 GHz
SENS:FREQ? - response 10GHZ - returns the current frequency
[SENSe]:BWIDth[:RESolution]:RATio <numeric_value>
[SENSe]BWIDth[:RESolution]:TYPE NORMal | FFT
[SENSe]:FREQuency:CENTer <numeric_value>
[SENSe]:FREQuency:STARt <numeric_value>
[SENSe]:FREQuency:STOp <numeric_value>
[SENSe]:FREQuency:TRACk <Boolean>
[SENSe]:FREQuency:VERify[:STATe] <Boolean>
[SENSe]:POWer:RLEVel:VERify:TOLerance <numeric_value>
[SENSe]:LIST:RANGe<1..20>:BWIDth[:RESolution] <numeric_value>
[SENSe]:LIST:RANGe<1..20>:FILTer:TYPE NORMal | FFT
[SENSe]:LIST:RANGe<1..20>:SWEep:COUNt <numeric_value>
[SENSe]:POWer:RLEVel:VERify[:STATe] <Boolean>
[SENSe]:POWer:RLEVel:VERify:TOLerance <numeric_value>
[SENSe]: POWer:TRACk <Boolean>
[SENSe]:SWEep:COUNt <numeric_value>
[SENSe]:SWEep:FORWard <Boolean>
[SENSe]:SWEep:MODe FAST | NORMal | AVERaged | MANual
[SENSe]:SWEep:POINts
[SENSe]:SWEep:TIME <numeric_value>
[SENSe]:BWIDth[:RESolution]:RATio <numeric_value>
The remote control command is used to specify the RBW value to be used for
each sub span as a ratio of the start frequency of the sub span. A value of 10%
specifies that the RBW should be set to 10% of the start frequency value. If the
required RBW value is no valid then the nearest RBW value will be set.
Example
SENS:BWID:RAT 1PCT
'Sets the RBW ratio to 1% of the start frequency
Characteristics
*RST value: 10 PCT
SCPI: conforming
[SENSe]BWIDth[:RESolution]:TYPE NORMal | FFT
The remote control command is used to specify whether to use the FFT Resolution
Filter or the conventional filter for each decade