Specifications

R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 86
INITiate:DISPlay <Boolean>
This command configures the behaviour of the display during a single sweep.
INITiate:DISPlay OFF
means that the display is switched off during the
measurement,
INITiate:DISPlay ON
means that the display is switched on
during the measurement.
The numeric suffix of INITiate is irrelevant with this command.
Example
INIT:DISP OFF
'The display is switched off during single measurements
Characteristics
*RST value: -
SCPI: device-specific
3.9 INPut Subsystem
The INPut subsystem controls the input characteristics of the RF inputs of the
instrument.
INPut:GAIN:STATe <Boolean>
INPut:PRESelection[:STATe] <Boolean>
INPut:GAIN:STATe <Boolean>
This remote control command switches the electronic 20 dB RF preamplifier into
the signal path (state ON) or removes it from the signal path (state OFF).
Example
INP:GAIN:STAT ON
'Switches the preamplifier into the signal path
Characteristics
*RST value: OFF
SCPI: device-specific
The command is only available with option B2 or B25.
INPut:PRESelection[:STATe] <Boolean>
This remote control command switches the preselection on or off.
Example
INP:PRES:STAT ON
'Preselection is switched on.
Characteristics
*RST value: OFF
SCPI: device-specific
The command is only available with the preselector option B2.