Specifications
R
&S
F
S-K40 Remote Control
Software Manual 1161.8209.42 - 04 85
FORMat[:DATA]:DEXPort:DSEParator POINt | COMMa
This command defines which decimal separator (decimal point or comma) is to be
used for outputting measurement data to the file in ASCII format. Different
languages of evaluation programs (eg MS-Excel) can thus be supported.
Example
FORM:DEXP:DSEP POIN
'Sets the decimal point as separator.
Characteristics
*RST value: POINt
SCPI: device-specific
3.8 INITiate Subsystem
The INITiate subsystem configures the instrument prior to a measurement being carried
out. It is basically used to tell the instrument which measurement is to be performed and
takes any necessary step to set up the instrument for the measurement.
INITiate[:IMMediate]
INITiate:CONTinuous <Boolean>
INITiate[:IMMediate]
This remote control command requests the R&S FS-K40 option to start a new
measurement sequence. If a measurement sequence is already in progress, then
the command will be ignored.
Example
INIT
'The R&S FS-K40 option will attempt to start a new measurement.
Characteristics
*RST value: -
SCPI: conforming
INITiate:CONTinuous <Boolean>
This command determines whether the trigger system is continuously initiated
(continuous) or performs single measurements (single).
Example
INIT:CONT OFF
'The R&S FS-K40 option will attempt a non continuous
measurement when initiated.
Characteristics
*RST value: -
SCPI: conforming