Technical data

6-8 Installation and Service Guide N5250-90001
Performance Tests and Checks PNA Series Microwave Network Analyzer System
System Performance Verification N5250A
Figure 6-3. System Verification Dialog Box
3. In the Calibration Kit box, select the “Agilent 85059A Data-based (1.00 mm)” calibration kit by clicking
on it. The corresponding verification kit to use is selected for you and displayed in the Verification Kit
box. Refer to Figure 6-3..
4. Under Outputs, the recommended selections are “File Tabular Data” and “File Graphs.” Other
selections may be made.
5. Click Run.
6. Enter the serial number for the 85059A kit and click Continue.
7. Install the 1.0 mm female to female cable (8.8 cm, Agilent part number 11500-60001) onto the right test
head (port 2) prior to calibration as shown in Figure 6-2. This cable is part of the 85059A 1.0 mm Precision
Calibration and Verification Kit and is considered to be a test port cable.
CAUTION Do not remove this test port cable once the Calibration/Verification process has begun. If
the test port cable becomes loose or is removed during the calibration/verification process,
the calibration is invalid.
CAUTION Do not pull on the connectors on the ends of the test port cable—this will damage the cable.