Technical data

Installation and Service Guide N5250-90001 4-5
PNA Series Microwave Network Analyzer System System Specifications
N5250A Specifications
N5250C Option 017 and 018 Test Head Bias-Tees Specifications
a. Pulse input connectors are operational only with Option H08 (Pulsed
Measurement Capability) enabled.
b. Based on deviation from signal reduction equation: Signal Reduction
(dB) = 20log
10
(Duty_cycle) = 20log
10
(pulse_width/period). Measured at
Pulse Repetition Frequency (PFR) of 1 MHz.
c. For rear panel LO port frequency, divide by 8.
d. For rear panel RF port frequency, divide by 6.
Table 4-3 Test Head Bias Input Connector
GNDU:
Sub Mini-Triaxial Connector, Trompeter BJ152 Insulated Bulkhead Jack (150
Series)
Sense:
Sub Mini-Triaxial Connector, Trompeter BJ152 Insulated Bulkhead Jack (150
Series)
Force:
Sub Mini-Triaxial Connector, Trompeter BJ152 Insulated Bulkhead Jack (150
Series)
Maximum Voltage: ±
30 VDC (typical)
Damage Voltage: ± 40 VDC
Maximum Current: ± 0.5 AMP