User manual
UDM 35/40 PANEL METER USER MANUAL
8
Scaling the Inputs
When is displayed, access is available for configuring the
electrical input range, the decimal point position and display span as
follows:
Electrical Input Range (HiE and LoE)
Allows the operator to define an electrical input range different from the
standard range setup at the “inP” programming step. As example, a UDM
with a full-scale 20mA (19.99mA) setting, it is possible to select an
electrical input range from 4.00mA (Lo.E) to 19.99mA (Hi.E) by
proceeding as follows:
When is displayed, press . “LoE” will appear.
Using the key, toggle to select a digit, then press the or key to
choose the desired value. Using the key, repeat as necessary until all
digits have proper values. Press .
“HiE” will now appear.
Using the same procedure as above, toggle through the digits using the
key and choose the desired values with the or key. Press to
confirm and advance to the next parameter, Decimal Point Position.