User`s guide
6-4 2G8647 2010-11-04
Specifications
Environmental Effects
• Heat Dissipation: total heat dissipation for the Imager is 4,800
BTU/hour
• Acoustical Noise:
– Less than 55 dB at 1.5 m during idle or standby
– Less than 65 dB momentary at 1 m during normal operation
Power Requirements
Four power cord sets are provided with this equipment:
• Power cord with a plug for use in North America
• Power cord with a plug for use in China
• Power cord with a plug for use in the United Kingdom
• Power cord with a plug for use in Europe
All other countries must use an Agency-approved power cord with plug
type suitable for the country of use, or contact an authorized Carestream
Health dealer.
One of the following single-phase, 15 A power sources, with grounding,
must be provided within 2.5 m (8 ft) of the Imager.
•100 Vac
±10%, 50/60 Hz ±3Hz
•120 Vac
±10%, 60 Hz ±3Hz
•200 Vac
±10%, 50/60 Hz ±3Hz
•220 Vac
±10%, 50/60 Hz ±3Hz
•240 Vac
±10%, 50/60 Hz ±3Hz
The wire must be insulation-rated for 600 Vac. A dedicated line is
recommended.
Network Requirements
The 6850 Laser Imager receives digital images from medical imaging
devices (modalities) over a 10/100Base-T or 1000Base-T Ethernet
Network. For 10/100Base-T Ethernet Networks, two CAT 5 UTP cables
are provided in the accessories package for the Imager; 2 m (6.5 ft) and
10 m (32.8 ft). For 1000Base-T Ethernet Networks, the user must
provide a CAT 6 UTP cable of sufficient length.