Specifications
Other Location
Considerations
Additional factors that influence where the Imager is located include
proximity to patients, flammable materials, liquids, and other equipment.
CAUTION:
This equipment is not contained in a sealed cabinet. Do not
use this equipment in locations where it can come in contact
with liquids, including body fluids.
IMPORTANT:
This equipment is intended to connect to other medical devices.
Only an authorized Service Provider of Carestream Health products
or Customers Qualified Service Personnel may install this
equipment.
Do not locate the Imager within 2 m (6.56 ft) of where a cell phone
will be used, even if the cell phone is separated from the equipment
by a wall.
Do not locate the Imager within 4 m (13.12 ft) of a microwave oven.
Electromagnetic radiation from a microwave oven is only an issue
if, after the oven door is closed and latched, the seal does not
maintain an electromagnetic tight fit between the oven door and the
oven main housing. Determining if the seal has an electromagnetic
tight fit requires special detection equipment.
Do not locate the Imager in the presence of flammable anesthetics,
oxygen, or nitrous oxide. The Imager does not have a gas-sealed
electronics enclosure and could ignite any flammable or explosive
gases present in the Imagers environment.
Do not locate the Imager in close contact with MRI devices due to
possible high magnetic fields near an MRI unit. The magnetic field
in the area where the Imager is installed must be less than 50 Gauss.
Do not substitute or modify any part of the Imager without prior
written approval of Carestream Health, Inc.