User`s guide

4-6 2G8647 2010-11-04
Adding a Modality and Configuring Modality Image Quality
Performing a Contrast
Test
There are three types of contrast test that can be run.
Standard - this test prints all TFTs from within the selected TFT Set.
IMPORTANT: The Custom and Broad sample tests use the TFT Set and
Contrast Table selected by you as a starting point. They
print in the middle of a sheet with nine images. The other
eight images on the test may use image processing from
any TFT Set/Contrast Table.
Custom - this test prints images whose image processing is close to
the starting point.
Broad sample - this test prints images whose image processing
varies more broadly from the starting point.
IMPORTANT: A contrast test always uses the last image sent by the
modality. You must disable printing before you send the
image you want to use for the contrast test.
1. From the Main Screen select, Utilities>Service>Printing enabled.
This will change the state to Printing disabled.
2. Select Configuration>Configure Modality.
3. Select the modality that will be used to run the contrast test.
4. Touch Select.
5. On the line for Maximum image density, select the check box for
Override Modality?
6. Touch Save.
7. Send an image from the modality to the Imager.
8. From the Modality Configuration screen, select the Contrast tab.
9. Select a Contrast Table and TFT Set to use as a starting point.
10. Touch Contrast Test.
11. Select the kind of test you want to run.
12. Touch:
Print
OK
Main
13. Enable printing. Select, Utilities>Service>Printing disabled.
This will change the state to Printing enabled.
14. Evaluate the test image.
15. If necessary, adjust the Contrast Table and TFT Set selections.
16. Adjust the setting for Maximum image density as appropriate.