Specifications
7 
CO
ppm
Provided the subject is not too distressed after seeing the breath CO 
measurement, then the calculated foetal carboxyhemaglobin level 
may be displayed by pushing the slide switch to the top position 
(FCOHb): 
FCOHb
%
The unit must not be used, however, for one second after the unit is 
first switched on i.e. before the display shows ‘BLOW’. Expired 
alveolar gas is then entrapped between sensor and mouthpiece 
valve. The display reading will rise to a plateau over the course of 
several seconds. The final value will be held until the unit is turned 
off and represents parts per million CO or %FCOHb depending on the 
slide switch position. 
The lights will come on according to the table on page 3.
Important note: 
Before repeating a measurement the unit must be turned off, 
and the mouthpiece and adapter removed for at least 1 
minute. This is to allow re-equilibration with ambient air and 
to dry the surface of the sensor. Visually inspect that all 
moisture has evaporated from the surface of the sensor 
before reuse. 