Specifications
6 
Operation 
For accurate results the BabyCO meter should be used at room 
temperature. If the instrument has been stored in cool or hot 
conditions then allow time to reach room temperature prior to use. 
Install the PP3 battery by sliding open the battery cover, clipping the 
battery in place and replacing the cover. Insert the mouthpiece 
adapter into the BabyCO meter and then insert a disposable 
mouthpiece into the adapter. Turn the unit on by pushing the slide 
switch to the middle position (CO). The display will show the 
software version number momentarily whilst correction for ambient 
levels is executed. During this time the unit must not be exposed to 
any source of CO. The buzzer will then sound and the display will 
change to: 
BLOW
Upon hearing the buzzer instruct the subject to inspire fully and hold 
their breath for as long as is comfortable. The breath-holding period 
is recommended to allow time for equilibration of alveolar gas. 
The subject should then seal their lips around the mouthpiece and 
exhale slowly and fully. 
After emptying the lungs completely the display will show the 
expired CO concentration in the expired air in parts per million 
(ppm): 