The MicroBlender provides selection of oxygen concentrations by means of a
single control knob located on the front of the unit. Oxygen concentrations
ranging from 21 to 100% are available.
Outlet Flow Range Bleed Flow
Low Flow
MicroBlender
Primary, Left Side 3-30 LPM No Bleed Flow
Auxiliary, Right Side 0-30 LPM 2.5-3.5 LPM
High Flow
MicroBlender
Primary, Bottom 15-120 LPM No Bleed Flow
Auxiliary, Right Side 2-100 LPM 10-12 LPM
Low Flow MicroBlender Outlets
High Flow MicroBlender Outlets
Primary
Outlet
Auxiliary
Outlet
Primary
Outlet
Auxiliary
Outlet