Instruction manual

Instruction Manual MicroBlender
L2965–101 Rev. D 21
Alarm Sound Generator Low Flow Reed Alarm
High Flow
Alarm Sound Intensity Low Flow 80 dB minimum at 1 foot
High Flow 80 dB minimum at 1 foot
Alarm/Bypass Reset Low Flow When inlet gas pressure
differential is 10 PSI or less
High Flow When inlet gas pressure
differential is 6 PSI or less
Pressure Drop Low Flow Less than 6 PSI at 50 PSIG
inlet pressures and 10 LPM
flow rate
High Flow Less than 6 PSI at 50 PSIG
inlet pressures and 40 LPM
flow rate
Weight Low Flow 2.75 lb. (1 .25kg)
High Flow
Di me nsion s (Excluding
Fittings)
Low Flow Height: 3 1/2 in. (8.9cm)
Width: 2 1/4 in. (5.8cm)
Depth: 3 5/8 in. (9.2cm)
High Flow Height: 3 1/2 in. (8.9cm)
Width: 2 1/4 in. (5.8cm)
Depth: 4 1/2 in. (11.5cm)
Note: Product specifications are subject to change without notice.