Section 10: Explanation of Abbreviations
Air/O
2
Mixture of Compressed Air and Oxygen
°C Degrees Centigrade
CGA Compressed Gas Association
DISS Diameter Indexed Safety System
°F Degrees Fahrenheit
FiO
2
Fractional Concentration of Inspired Oxygen
O
2
Oxygen
LPM Liter Per Minute
P/N Part Number
PSIG Pounds Per Square Inch Gauge
Do not operate the MicroBlender outside the supply pressure range (30–75
PSIG). Gas supply pressures of 50 PSIG provide optimal performance. The graphs
on the following page illustrate typical flow performance characteristics of a
representative sample for the Low Flow MicroBlender. The graphs are typical
of a representative sample; slight variations among units should be expected.