Instruction manual

MicroBlender Instruction Manual
8 L2965–101 Rev. D
Warning!
If either the air or oxygen gas source fails, the MicroBlender
alarm sounds, alerting the clinician that a condition has occurred
that may significantly alter the FiO
2
and flow output from the
MicroBlender.
If either the air or oxygen gas source pressure is reduced or
increased creating a pressure differential of 20 ±2 PSI, the
MicroBlender alarm sounds. This condition significantly alters the
FiO
2
and flow output from the MicroBlender.
Always operate the MicroBlender with clean and dry medical
grade gases.
Air Inlet Filter/Water Trap (P/N 07426) is recommended for use
with the MicroBlender.
The patient gas must be monitored with an oxygen analyzer.
DO NOT steam clean, autoclave, or otherwise subject the
MicroBlender to temperatures above 145°F (62°C).
DO NOT immerse the assembled MicroBlender in liquid
decontamination agents.
Consult a physician for appropriate FiO
2
setting.
DO NOT tape, obstruct, or remove the reed alarm outlet at any
time.
DO NOT occlude or obstruct the bleed port or muffler on the
bottom of the MicroBlender.
Adjustment of the oxygen concentration must be verified using
an oxygen analyzer.