Service manual

80 Infant Flow
®
SiPAP
675-120 Revision D May 2005
10. Accept the default BiPhasic settings by selecting the BiPhasic soft key again (Figure 4).
Figure 4
11. Verify that the measured value doesn’t exceed 23% FiO2. If so, then troubleshoot for an
oxygen leak. Check all the fittings and connections and repeat the Oxygen Leak Test
(Figure 5).
Figure 5