Service manual

Service Manual 73
675-120 Revision D May 2005
10 Click on the Download button. The download takes approximately 1 minute at
57600 bps. When the download is complete, the audible alarm on the IFSD stops
and the unit enters the normal operational mode.
11 Disconnect the Download Cable from plug PL4 on the Main Processor PCB.
12 Refit the case.
13 Carry out a functional check of the IFSD.