Service manual

64 Infant Flow
®
SiPAP
675-120 Revision D May 2005
Removal and Fitting of Oxygen Filter
Note
This procedure can be done without removing the case.
Refer to Figure 14.
Removal
1 Remove the banjo screw (1) on the end of the air/oxygen mounting block (2).
2 Remove the filter (3).
3 Remove and discard the O-ring (4).
Fitting
1 Install a new O-ring (4) on the banjo screw (1).
2 Install a new filter (3) in the air/oxygen mounting block and install the banjo
screw (1).
3 Tighten the banjo screw (1).
Figure 14 - Oxygen Filter Removal/Fitting