Service manual

62 Infant Flow
®
SiPAP
675-120 Revision D May 2005
Removal and Fitting of Case
Refer to Figure 12.
Removal
1 Remove the ten screws (1) on the rear panel.
2 Pull off the case (2).
Fitting
1 Install the case (2) over the main frame, making sure that all wires and pipes
are secure and cannot be trapped by the case.
2 Install the ten screws (1) on the rear panel and tighten.
3 Carry out an Oxygen Leak check (see Appendix A - Oxygen Leak Test) and
functional check of the IFSD.
Note: An Oxygen Leak test must be done after every service that requires removal of
the case to ensure no buildup of oxygen is present within the case.
Figure 12 - Case Removal/Fitting
Removal and Fitting of Battery
Refer to Figure 13.
Removal
1 Remove the case.