Chapter 7- Maintenance
Cleaning
CAUTION
Do not immerse any part of the IFSD in water or sterilize it with gas or steam.
1 Clean the exterior surfaces of the IFSD and the transducer interface with a mild
soap or liquid disinfectant solution. Do not use cleaning agents that contain
abrasives.
2 Make sure that cleaning agents do not enter the unit through patient
connection ports.
Maintenance
WARNING
Oxygen vigorously accelerates combustion. To avoid explosion hazard, do not
use any instrument or other equipment that may have been exposed to oil or
grease contamination.
General
1 Examine the exterior of the case for damage and dirt. If necessary clean the
unit.
2 Check the water trap on the back panel. If water is visible in the water trap,
push the button on the bottom of the water trap to release the water.
3 Technical Assistance can be obtained from the offices listed at the beginning
of this manual.
4 Store the unit in a clean dry location. Make sure that the connections and ports
are suitably blanked to prevent the ingress of dirt, moisture and foreign objects. If
the unit is not being used for a long period of time, remove the battery.
5 Dispose of scrap units in accordance with the local regulations
Removal and Fitting of Parts – General
1 Only parts approved by VIASYS Healthcare may be used in this unit.
2 Remove all attached patient connections before removal of parts. Disconnect
the unit from the power supply.
3 Place the unit on a stable, clean work surface before removal of parts.