Service manual
48 Infant Flow
®
SiPAP
675-120 Revision D May 2005
Alarm Method of Setting
Indications and
Actions
Method of Resetting
or Canceling
Low pressure
(<
2 cmH
2
O below set
point for 15 seconds)
or
<
1 cmH
2
O at any time).
Set automatically on
entering NCPAP
mode for the first time
after start up or by
pushing the alarm
button for three
seconds at any time.
Intermittent high level
Audible Alarm.
Warning Bar flashes.
Pressure display
flashes.
Digital pressure low
limit is highlighted.
Alarm Button flashes.
Push the Alarm
Button once to stop
the Audible Alarm for
30 seconds (Alarm
Button flashes, digital
pressure low limit
stays highlighted,
Warning Bar still
flashes ).
Push the Alarm
Button for three
seconds to reset the
limit (alarms clear).
Over pressure (Patient
pressure
> 11 cmH
2
O when in
BiPhasic or BiPhasic Tr
mode).
Always active when
power is on.
Intermittent high level
Audible Alarm.
Warning Bar flashes.
O
2
digital display
flashes.
Digital O
2
alarm low
limit is highlighted.
Alarm Button flashes.
Reduce pressure to
below the high
pressure limit then
push the Alarm
Button for three
seconds.
High BiPhasic/ BiPhasic
Tr pressure
(MAP >
3 cmH
2
O above
set point at time alarm is
set for 15 seconds).
Set automatically on
entering NCPAP
mode for the first time
after start up or by
pushing the alarm
button for three
seconds at any time.
Intermittent high level
Audible Alarm.
Warning Bar flashes.
High MAP pressure
display flashes.
Digital high MAP
pressure limit is
highlighted.
Alarm Button flashes.
Push the Alarm
Button once to stop
the Audible Alarm for
30 seconds (Alarm
Button flashes, digital
high MAP pressure
limit stays
highlighted, Warning
Bar still flashes).
Push the Alarm
Button for three
seconds to reset the
limit (alarms clear).
Low battery charge
(<
40%).
Automatic Battery status indicator
changes from gray to
red.
Connect external
power.