Service manual

44 Infant Flow
®
SiPAP
675-120 Revision D May 2005
Confirm the BiPhasic Tr settings by touching the flashing BiPhasic Tr button.
When the settings have been accepted, the screen changes to the locked screen
and treatment starts.
To adjust the parameters, touch the
button.
The screen changes to the BIPHASIC TR Adjust Screen.
Parameters Ti, Rb and Tapnea can now be adjusted.
To return to the Mode Selection Screen, touch the
button.
Note
The button is blank if an alarm condition occurs. If this occurs, the button
cannot be operated until the alarm condition has been cleared or silenced using the
alarm button.
After two minutes if the Operator has not made any inputs, the screen changes to the
Locked Screen.
the key pad is locked
the display remains as shown for patient monitoring
Touch the
button to unlock the key pad.
the display returns to the BiPhasic Tr Adjust Screen
Note
If a high priority alarm occurs, the key pad automatically unlocks.
Calibration
WARNING
Calibration must only be done when the unit is not connected to the patient.
From the Start Up Screen touch the button.
The screen changes to the Calibration Screen.
the alarm limits are disabled
Turn the FiO
2
control to 21 and confirm by touching the flashing button.
Turn the FiO
2
control to 100 and confirm by touching the flashing button.