Service manual
20 Infant Flow
®
SiPAP
675-120 Revision D May 2005
Cautions
• Federal Law (USA) restricts this device to sale by or on the order of a
physician.
• The precision gas blender in this product may become non-functional or
damaged if used without the protective water trap and filters provided.
• The power switch on this unit does not isolate the external power supply.
Disconnect the power supply cable to ensure complete isolation.
• Before use, verify that this equipment has been authorized for use by a
qualified person.
• The Infant Flow SiPAP™ has been designed and tested as a complete
system using Infant Flow™ accessories. Only accessories approved for use
should be used. If in doubt, please contact your local VIASYS representative.
• Where the integrity of the external protective earth conductor is in doubt, the
equipment shall be powered by its internal power source (battery).
• Do not immerse any part of the IFSD in water or sterilize it with gas or steam.
• Ensure patient breathing circuit is replaced at regular intervals.