Service manual

Service Manual 19
675-120 Revision D May 2005
When filling a humidifier, do not move the stand. Moving or transporting the
stand while refilling may cause the stand and equipment to over balance.
Do not use conductive patient circuits with the Infant Flow SiPAP Driver.
Nasal CPAP can cause nasal irritation, septal distortion, skin irritation and
pressure necrosis. Observe the usage guidelines to minimize these
complications.
This device exhausts O
2
during normal operation. Oxygen vigorously
accelerates combustion. To avoid fire hazard, do not place flammable
materials or sources of heat close to the exhaust.
Do not use the equipment without the exhaust tube fitted (refer to Figure 2).
To reduce trip hazard, always ensure cable and tubes are restrained away
from walking areas.
The Abdominal Respiratory Sensor will not detect all forms of apnea.
Independent monitoring should always be used with this device.
If the unit is shelf mounted, ensure that the unit is stable and that hoses and
cables are restrained to avoid hazard of toppling.
This equipment is not suitable for use in the presence of a flammable
anesthetic mixture.
The NCPAP Pres High flowmeter must be adjusted to zero when not required
for the patient.
Under extreme conditions (minimum supply pressure and maximum gas
demand, including auxiliary output) output flow rates and delivered pressure
may be reduced.
Only use the supplied AC cable to connect to the power supply.
Do not attach the Generator to the patient until the initial set up is complete.
The
indicates a connection between the transducer interface and the
unit. It does not indicate correct positioning of the Abdominal Respiratory
Sensor.
Calibration must only be done when the unit is not connected to the patient.
Ensure the whistle sounds during gas connection. If not, the device may be
faulty and should not be used. Refer to your Service Engineer.
Verify that the displayed value for delivered FiO
2
corresponds to the value set
on the blender. Refer to Faults and Indications.
Oxygen vigorously accelerates combustion. To avoid explosion hazard, do
not use any instrument or other equipment that may have been exposed to oil
or grease contamination.