Service manual

Service Manual 15
675-120 Revision D May 2005
Alarms
High airway pressure – 3 cmH20 above measured airway pressure
High circuit pressure – maximum 11 cmH20 in time triggered Biphasic mode
High circuit pressure – maximum 15 cmH20 in patient triggered Biphasic tr*
mode
Low airway pressure – 2 cmH20 below measured airway pressure or 1
cmH20 if otherwise would be zero
High and Low delivered Oxygen concentration +5% of setting
Low breath rate alarm
Low or Flat battery charge level
Input gases failure
Alarm volume (electronic alarms) 70 dBa at 1 meter
Pneumatic Supply
Patient Gas Outlet - 15 mm standard taper fitting
Patient Pressure Input - 4.5 mm Luer taper fitting
Gas Supply - Nominal 4 bar, clean, dry medical air and oxygen
Range - 2.8 - 6 bar; Maximum differential pressure 2 bar
Manometer - Range 0 to + 20 cmH2O, accuracy, ± 2% of span
Gas Connections - Standard DISS or NIST connectors
Electrical Supply
Input Voltage -100-230 VAC
Input Frequency -50/60 Hz
Power Consumption -50 VA maximum
Fuse Rating For 220 V nominal operation-“T” Type 2.5 A at 250 V
Device Housing Protection rating level -IPX1
Battery Working Time -2 hours (from fully charged state)
Battery Charging Time –max. 16 hours