Service manual
Service Manual 15
675-120 Revision D May 2005
Alarms
• High airway pressure – 3 cmH20 above measured airway pressure
• High circuit pressure – maximum 11 cmH20 in time triggered Biphasic mode
• High circuit pressure – maximum 15 cmH20 in patient triggered Biphasic tr*
mode
• Low airway pressure – 2 cmH20 below measured airway pressure or 1
cmH20 if otherwise would be zero
• High and Low delivered Oxygen concentration +5% of setting
• Low breath rate alarm
• Low or Flat battery charge level
• Input gases failure
• Alarm volume (electronic alarms) 70 dBa at 1 meter
Pneumatic Supply
• Patient Gas Outlet - 15 mm standard taper fitting
• Patient Pressure Input - 4.5 mm Luer taper fitting
• Gas Supply - Nominal 4 bar, clean, dry medical air and oxygen
• Range - 2.8 - 6 bar; Maximum differential pressure 2 bar
• Manometer - Range 0 to + 20 cmH2O, accuracy, ± 2% of span
• Gas Connections - Standard DISS or NIST connectors
Electrical Supply
• Input Voltage -100-230 VAC
• Input Frequency -50/60 Hz
• Power Consumption -50 VA maximum
• Fuse Rating For 220 V nominal operation-“T” Type 2.5 A at 250 V
• Device Housing Protection rating level -IPX1
• Battery Working Time -2 hours (from fully charged state)
• Battery Charging Time –max. 16 hours