Service manual
14 Infant Flow
®
SiPAP
675-120 Revision D May 2005
Chapter 2 - Product Specifications
Modes
• NCPAP
• NCPAP with breath rate monitoring and low rate alarm
• BiPhasic (time triggered)
• BiPhasic (time triggered) with breath rate monitoring and low rate alarm
• BiPhasic tr (patient triggered ) with breath rate monitoring, low breath rate
alarm, and apnea back up
Controls
• Inspiratory Time (Ti) – 0.1-3.0 seconds
• Rate (R) – 1-120 (Comprehensive* only)
• Rate (R) – 1-54 (Plus only)
• Apnea Interval (T
apnea) – 10-30 seconds, 5 second intervals (Comp* only)
• Apnea Interval (T
LBR) – 10-30 seconds, 5 second intervals (Plus only)
• NCPAP/Pres Low flow meter – 0-15L/min, accuracy +/- 15% of selected
output
• NCPAP/Pres High flow meter – 0-5L/min, accuracy +/- 15% of selected
output
• Manual Breath – X 1
• Rate monitoring on/off – NCPAP
• %O
2
– 21 - 100% - accuracy +/-3%
Monitors
• CPAP
• PEEP
• MAP
• PIP
• %O
2
• I:E ratio
• Spontaneous rate (Rs)
• Battery charge level