Specifications

16
3%450!.$#!,)"2!4)/.02/#%$52%CONT
3:?"2%!+2!.'%a
The break range is a number from 0 to 64 that corresponds to the number of graduations
of weight change to break out of filtering. Note that the higher the number,the greater the
filtering.
Observe the current setting. If the value displayed is acceptable, press the %.4%2 key to
save it. Otherwise, use the numeric keys to enter a new value, then press the %.4%2 key
to save it. Allowable values are: 0 through 64.
+:?3!-0,%2!4%
The Sample rate value for filter number 3 is a value which represents the rate at which the
displayed weight is refreshed. Note, that the lower the number, the slower the refresh and
the more resistant the weight will be to vibration or other causes of weight deviation.
Observe the current setting. If the value displayed is acceptable, press the %.4%2 key to
save it. Otherwise, use the numeric keys to enter a new value, then press the %.4%2 key
to save it. Allowable values are 1 to 8 samples per second.
4HEFOLLOWINGPROMPTSAREFORPRINTEROUTPUTONLY
(?a3%2)!,/54054"!5$2!4%
With the display showing 32(4?, observe the current setting. If the value displayed is
acceptable, press the %.4%2 key to save it. Otherwise, use the numeric keys to enter a new
value, then press the %.4%2 key to save it. Allowable values are: 12, 24, 48, 96, 19, 38.
12 = 1200 Baud 24 = 2400 Baud 48 = 4800 Baud
96 = 9600 Baud 19 = 19.2k Baud 38 = 38.4k Baud
%&?a0!2)493%44).'
With the display showing %2:?, observe the current setting. If the value displayed is
acceptable, press the %.4%2 key to save it. Otherwise, use the numeric keys to enter a new
value, then press the %.4%2 key to save it. Allowable values are: 0, 1 or 2.
0 = No parity (8-N-1) 1 = Even parity (7-E-1) 2 = Odd parity (7-O-1)
%&'?a,!"%,02).4%23%,%#4)/.
With the display showing %:<?, observe the current setting. If the value displayed is
acceptable, press the %.4%2 key to save it. Otherwise, use the numeric keys to enter a new
value, then press the %.4%2 key to save it. Allowable values are: 0, 1, 2, 3 or 4.
0 = P200 (Eltron Companion Plus)
P220 / P240 (Eltron 2722 / 2742)
1 = user defined (via “n-Control”)
2 = P200, P205, or P220 / P240 with DSR handshake check
3 = user defined (via “n-Control”) with DSR handshake check
4 = P205 (Cognitive Del-Sol)