SETUP AND CALIBRATION, CONT.
dPP= (Decimal Point Position)
With the display showing dpp=, press the ZERO key to show the current setting. If the value
displayed is acceptable, press the ZERO key again to save it. Otherwise, use the UNITS key
to step through the selections for a new value. When the desired value is displayed, press the
ZERO key to save it. Allowable values are:
1= XXXXX.X 2 = XXXX.XX 3 = XXX.XXX
CAP= (Scale Capacity)
With the display showing CAP=, press the ZERO key to show the current setting. If the value
displayed is acceptable, press the ZERO key again to save it. Otherwise, use the UNITS key
and the ON/OFF key to select the capacity. Note that the cursor location is identified by the
blinking character and can be advanced to the next position (left) by pressing the ON/OFF
key. Pressing the UNITS key will change the blinking character to the next value. Verify that
the numbers selected are the same as the weight listed on the capacity label and then press
the ZERO key to save it.
CAL= (Calibration)
With display showing CAL=, press ZERO key. The display will change to show current setting
0 (0 = NO, skip calibration). If scale has been previously calibrated and you wish to skip
calibration and proceed to Store Setup and Calibration, press ZERO key again.
To begin calibration, press UNITS key to select 1 (YES, calibrate scale) and then press ZERO
key. Display will change to LOAd=.
LOAd= (Load Calibration Weight)
The display will change to LOAd= which is a prompt for the entry of the calibration weight
value and placement of this amount of test weights on the scale platform.
1. Make certain the scale platform is empty and free of debris, then place the desired
amount of calibrated test weights on the scale platform. It is recommended that a
minimum of 50% of the scale's capacity be used but 70% to 100% is preferred.
2. Press the ZERO key.
3. Determine the exact amount of test weights to be placed on the scale platform and then
using the UNITS key and the ON/OFF key select this value. Note that the cursor location
is identified by the blinking character and can be advanced to the next position (left) by
pressing the ON/OFF key. Pressing the UNITS key will change the blinking character to
the next value. Verify that the numbers selected are the same as the total weight of the
test weights.
4. Press the ZERO key.
5. Starting at the left and proceeding right, a series of dashes will appear on the display. The
dashes will disappear and the display will change to UnLd.
UnLd (Unload Calibration Weight)
1. With display showing
UnLd, remove the test weights from the scale platform.
2. Press the ZERO key.
3. Starting at the left and proceeding right, a series of dashes will appear on the display. The
dashes will disappear and the display will change to off and the scale will turn off.
STORE SETUP AND CALIBRATION
To complete setup and calibration and store
setup and calibration data, perform the following:
1. Press and hold (approximately 2 seconds) ON/OFF key until scale is turned on.
2. The display will immediately change to off and the scale will turn off.
3. Setup and calibration data is now stored in scale's nonvolatile memory.
4. Press and hold (approximately 2 seconds) ON/OFF key until scale is turned on.
5. Scale is now ready for operation.