PZ3000 Series Ingredient Scale Owner’s Manual 8528-M372-O1 Rev E 05/14 CARDINAL SCALE MFG. CO. PO BOX 151 WEBB CITY, MO 64870 PH (417) 673-4631 FAX (417) 673-5001 www.detectoscale.com Technical Support: Ph: 866-254-8261 techsupport@cardet.
INTRODUCTION TABLE OF CONTENTS We wish to thank you for your purchase of our SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . 1 PZ3000 Ingredient scale. It has been designed INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . 3 and manufactured at our factory in Webb City, POWER CONNECTION . . . . . . . . . . . . . . . . . . 4 MO U.S.A. with quality and reliability. Power Supply . . . . . . . . . . . . . . . . . . . . . . . . .
Serial Number_______________________ Date of Purchase ____________________ Purchased Form_____________________ ___________________________________ ___________________________________ RETAIN THIS INFORMATION FOR FUTURE USE PRECAUTIONS Before using this instrument, read this manual and pay special attention to all "NOTIFICATION" symbols: IMPORTANT ELECTRICAL WARNING 2 8528-M372-O1 Rev E PZ3000 Series
INSTALLATION 1. Carefully remove the scale base and display from the shipping carton and inspect them for any evidence of damage (such as exterior dents and scratches) that may have taken place during shipment. Remove the packing material from the scale base and on the PZ3030, the shipping screw on the bottom of the scale base. Next, remove the protective covering from the display and keypad.
POWER CONNECTION NOTE: The PZ3000 can be operated from a power supply or from 6 “AA” size, (2300mAh minimum) NiMH batteries. The power supply is also used to recharge the NiMH batteries. Batteries are available by ordering Cardinal part number, 6800-0020. Power Supply To power the PZ3000 using the power supply, connect the power supply's connector into the power jack on the bottom side of the display and then connect the 12 VDC, 1 Amp power adapter into the proper electrical outlet. See Figure 1.
POWER CONNECTION, CONT. 7. Slide one side of the battery door so that it clears the edge of the indicator enclosure and then remove the battery door. 8. Place the battery door on a table or bench with the battery holder and battery strap up. 9. Referring to Figure No. 3, remove the 2 Philips screws securing the battery strap to the battery door and remove the strap. Battery Strap Battery Connector Figure No. 3 Remove Philips Screws 10.
POWER CONNECTION, CONT. Figure No. 5 - + + - - + + - - + + - 6 “AA” Size NiMH Batteries (2300mAh minimum) 15. Insuring no wires are between the battery door and the back of the indicator, lower the battery door. 16. Install the two screws and washers removed in step 4. NOTE: The PZ3015 and PZ3030 use thumbscrews and should be finger-tightened only. 17. Return the indicator to the scale, wall or table stand. 18.
POWER CONNECTION, CONT. 9. Referring to Figure No. 6, locate the battery charger jack on the battery door. 10. Plug the AC power adapter (battery charger) into the jack on the battery door and then connect the AC power adapter (battery charger) into the proper electrical outlet. 11. Allow approximately 4 to 8 hours to fully recharge the batteries. Battery Connector Figure No.
KEYPAD DESCRIPTION The PZ3000 display is equipped with a membrane keypad containing 3 keys and the LITE-WAY photo-eye sensor. The keypad is shown in detail in Figure 8. Figure No. 8 DO NOT operate the keypad with pointed objects (pencils, pens, etc). Damage to keypad resulting from this practice is NOT covered under warranty. The following describes the use of each key. Pressing this key will apply power to the instrument and turn on the scale.
ANNUNCIATORS The annunciators are turned on to indicate that the display is in the mode corresponding to the annunciator label or that the status indicated by the label is active. The low battery annunciator is located in the upper left corner of the display. It is used to indicate the battery status. Refer to the Power Connection section of this manual for more details. GROSS This annunciator is turned on to show that gross weight is displayed. Gross weight will be displayed when no tare weight is stored.
OPERATION 1. Press and hold (approximately 2 seconds) the ON/OFF key until the scale is powered up. 2. With no weight on the scale, press the ZERO key. 1 3. After a brief moment, the display will show 00 )0 4. The scale is now “zeroed” and ready for use. 5. Place container on the scale, the weight of the container will be displayed. 6. Press the ZERO key to return the scale to zero. 7. Add your first ingredient to the desired weight; re-zero the scale for the second ingredient by pressing the ZERO key. 8.
SETUP AND CALIBRATION Your PZ3000 Ingredient scale has been pre-configured at the factory and should not require configuration for use in most applications. In the event that the factory settings do not meet the requirements of your application, the following describes the steps to configure the scale. IMPORTANT! The display and scale base are a matched pair. Do not switch them (refer to S/N tags) without recalibration. To Begin Setup and Calibration: 1. Press and hold the ZERO key. 2.
SETUP AND CALIBRATION, CONT. FLT= (Digital Filter Level Selection) Your scale will arrive with factory filter settings of 0 = Minimal (PZ3060 will be 2 = Heavy). Please check with Tech Support before changing filter level, break range and sample rate. Press ZERO key to show current setting. If value displayed is acceptable, press ZERO key again to save it. Otherwise, use UNITS key to step through selections for a new value and then press ZERO key to save it.
SETUP AND CALIBRATION, CONT. dPP= (Decimal Point Position) With the display showing dpp=, press the ZERO key to show the current setting. If the value displayed is acceptable, press the ZERO key again to save it. Otherwise, use the UNITS key to step through the selections for a new value. When the desired value is displayed, press the ZERO key to save it. Allowable values are: 1= XXXXX.X 2 = XXXX.XX 3 = XXX.
TROUBLESHOOTING The following describes conditions you may encounter while operating this scale. Instructions are provide to help you correct the problem. If you are still unable to resume normal operating functions, please contact your nearest scale technician or our Technical Support Department. Negative Weight Display -9 15 The scale will display a negative weight when the scale base has been emptied or partially emptied without re-zeroing the scale.
PARTS IDENTIFICATION (Display Assembly) ITEM QTY PART NUMBER DESCRIPTION 1 1 593GR986 SERIAL TAG 2 4 6013-0039 NUT HEX #6-32 3 2 6021-0661 SCW PAN HEAD #6-32 X .25 X S.S. 4 2 6021-1032 THUMB SCRW, 6-32 X 0.25 (PZ3015, 3015L & 3030) 6021-0300 SCW SOCKET HEAD CAP SCREW (PZ3060) 5 2 6610-2000 JACK SOCKET 6 6 6680-0004 WASHER LOCK INT. TOOTH #6 Z/P 7 4 6680-0045 SPACER (PCB) #6 X .
PARTS IDENTIFICATION (Battery Door Assembly) ITEM QTY PART NUMBER DESCRIPTION 1 3 6013-0039 NUT HEX #6-32 2 2 6021-0654 SCW PAN HEAD #6-32 x .250 PDMS 3 2 6021-0687 SCW TRUSS HEAD #6-32 x .312 THMS 4 3 6680-0004 WASHER LOCK INT. TOOTH #6 Z/P 5 1 8528-B272-08 LABEL: BATTERY REQ.
PARTS IDENTIFICATION (PZ3015 Base Assembly) 8528-M372-O1 Rev E PZ3000 Series 17
PARTS IDENTIFICATION (PZ3015 Base Assembly) ITEM QTY PART NUMBER 1 1 428R1079 DECAL 2 1 427R1366 COMMODITY TRAY 3 1 427R1365 WEIGHBRIDGE 4 1 8528-B110-0A 5 2 587R9378 LOAD CELL SPACER 6 2 6021-1124 #10-32 X .75 SOC HD S.S. 7 1 6021-1258 #10-32 X .375 SET SCW 8 2 6021-1074 #10-32 X .5 FHMS S.S.
PARTS IDENTIFICATION (PZ3015L, PZ3030 Base Assembly) 8528-M372-O1 Rev E PZ3000 Series 19
PARTS IDENTIFICATION (PZ3015L, PZ3030 Base Assembly) ITEM QTY PART NUMBER 1 1 8528-C262-0A BASE PLATE WELDMENT 2 1 8528-D264-08 WEIGHBRIDGE 3 1 8528-D263-08 COMMODITY TRAY 4 1 8528-B110-0A LOAD CELL ASSEMBLY DF-50T 5 2 6021-2003 FLAT HEAD SCREW .
PARTS IDENTIFICATION (PZ3060 Base Assembly) 8528-M372-O1 Rev E PZ3000 Series 21
PARTS IDENTIFICATION (PZ3060 Base Assembly) ITEM QTY PART NUMBER 1 1 1961-B006-08 LABEL, DETECTO 2 1 1961-D002-08 BASE PLATE 3 1 1961-D004-08 WEIGHBRIDGE 4 1 1961-D005-08 COMMODITY TRAY 5 1 2950-C121-4A LOAD CELL TSP-100KG 6 1 593GR986 SERIAL TAG ASSEMBLY 7 2 6007-0009 HEX HEAD BOLT 1/4-20 X 1/2” Z.P. 8 4 6007-0011 BOLT, HEX HEAD 1/4-20 X 1” Z.P. 9 10 6013-0045 NUT, HEX 1/4-20 10 4 6021-1013 #10-32 X .375 RHMS 11 4 6021-1429 1/4-20 X 0.
NOTES 8528-M372-O1 Rev E PZ3000 Series 23
STATEMENT OF LIMITED WARRANTY Detecto Scale warrants its equipment to be free from defects in material and workmanship as follows: Detecto warrants to the original purchaser only that it will repair or replace any part of equipment which is defective in material or workmanship for a period of one (1) year from date of shipment. Detecto shall be the sole judge of what constitutes a defect.
STATEMENT OF LIMITED WARRANTY Conditions Which Void Limited Warranty This warranty shall not apply to equipment which: A.) Has been tampered with, defaced, mishandled or have had repairs and modifications not authorized by Detecto. B.) Has had serial number altered, defaced, or removed. C.) Has not been grounded according to Detecto’s recommended procedure. Freight Carrier Damage Claims for equipment damaged in transit must be referred to the freight carrier in accordance with freight carrier regulations.