ProDoc Series Digital Physician Scale PD100 PD200 PD300 PD300DHR (Digital Height Rod) and PD300MHR (Mechanical Height Rod) Owner’s Manual 0044-M340-O1 Rev A 05/08 QuickMedical - Authorized Distributor PO Box 1052 y Snoqualmie, WA 98065 Ph:888-345-4858 y Fax: 421-831-6032 www.quickmedical.com/detecto/ Technical Support: 0044-M340-O1 y ProDoc Series Ph: 866-254-8261 y techsupport@cardet.
0044-M340-O1 y ProDoc Series Page 2
TABLE OF CONTENTS SPECIFICATIONS ………………………………………………………………………………. Standard Features …..…………………………………………………………………….… Care and Cleaning …..…………………………………………………………………….… PD100 …………………………………………………………………………………………….. Unpacking ……………………………………………………………………………………. Assembly ……………………………………………………………………………………... Quick Start ……………………………………………………………………………………. Placing the Scale …………………………………………………………………………..… Placing the Display ….……………………………………………………………………..… Wall Mounting …………………………………………………………………………….
INTRODUCTION Thank you for purchasing our Detecto ProDoc Series Digital Physician Scale. Your scale has been designed for simple and straightforward use and to assure accuracy and dependability for years to come. Please read this manual thoroughly before attempting to use your scale and keep it handy for future reference. It contains important instructions for installation and proper operation of your scale.
SPECIFICATIONS Model Number ………………… PD100 (Floor Scale with Remote Display) PD200 (Floor Scale PD300 (Eye-Level Scale) PD300DHR (Eye-Level Scale with Digital Height Rod) PD300MHR (Eye-Level Scale with Mechanical Height Rod) Capacity ...……………………….. 480 lb x 0.2 lb (220 kg x 0.1 kg) Weight Units ......
This page intentionally left blank.
PD100 Unpacking Remove the scale base, display, hardware pack and other components from the shipping carton. After removing from the carton, check for any damage that may have taken place during shipment. Keep and use the original carton and packing material for return shipment if it should become necessary. The purchaser is responsible for filing all claims for any damages or loss incurred during transit.
PD100 Placing the Scale The scale should be placed on a flat, level floor or low cut carpet away from any rapidly moving air source (heating and cooling vents). Check to make certain the scale is level. The level indicator is located at the rear of the scale. If the scale is not level (the bubble will not be centered), loosen the locking ring on all four (4) feet and adjust them as required to center the bubble and attain a level scale.
PD100 Placing the Display Wall Mounting, Cont. 7. Referring to Figure No. 2, make sure the four (4) feet of Wall Bracket “A” are placed in the bracket. Display Feet (for Desk or Table Mounting) Wall Bracket “A” Figure No. 2 Figure No. 3 8. Next, locate the four (4) slots on the back of the display and align them with the four (4) tabs on the Wall Bracket “A”. Refer to Figure No. 3. 9. Insert the tabs into the slots and press together until the bracket locks to the back of the display. 10.
PD100 Placing the Display Desk or Table Mounting 1. The location chosen should be a stable, level surface (either a desk or solid table), free of temperature extremes and water. The display should be where it can be easily viewed, not subject to direct sunlight and where the keypad is within easy reach of the operator. 2. Referring to Figure No. 5, remove the four (4) feet from inside Wall Bracket “A” and install them in each corner of the bracket as shown in Figure No. 6.
PD200 Unpacking Remove the scale from the carton by grasping the column (not the display) and lifting up with equal force on the column and the scale base. After removing from the carton, check for any damage that may have taken place during shipment. Keep and use the original carton and packing material for return shipment if it should become necessary. The purchaser is responsible for filing all claims for any damages or loss incurred during transit.
PD300 Unpacking Remove the scale base, display, hardware pack and other components from the shipping carton. After removing from the carton, check for any damage that may have taken place during shipment. Keep and use the original carton and packing material for return shipment if it should become necessary. The purchaser is responsible for filing all claims for any damages or loss incurred during transit.
PD300 Assembly Installing the Column and Display, Cont. 2. Insert the socket end of the connection cable into the column connector case. Feed cable through entire length of column. Refer to Figure No. 10 to insure correct connector on each end of column. Column Connector Case Connection Cable Figure No. 10 3. Place the scale on a desk or table Connection Cable 4. Connect the socket end of the connection cable to the preset cable in the scale base. Preset Cable Figure No. 11 5.
PD300 Assembly Installing the Column and Display, Cont. 6. Insert the M4 machine screw through the column holder into the column and tighten the screw to attach the column to the column holder. Column Holder Column M4 Machine Screw Figure No.13 7. Slide the column cover down the column and into the column holder. Column Cover 7.1. Locate the two (2) tabs on the front of the column cover and the two (2) slots in the column holder. Column Holder 7.2.
PD300 Assembly Installing the Column and Display, Cont. Display 9. Remove the scale from the desk or table and place it upright on the floor. Preset Cable 10. Connect the plug end of the connection cable to the preset cable from the display and then insert the cable into the column. Tab Connection Cable 11. Locate the two (2) tabs (one on each side) on the connector case and the two (2) slots in the bottom of the display. 11.1.
PD300 Quick Start Step 1. Once the scale is assembled, place it on the floor. Step 2. Carefully lay the scale over on its side with the edge of the display resting on the floor and then locate the battery access cover on the bottom of the scale. Step 3. Remove the cover, install six (6) “AA” size batteries1 and then replace the cover. Step 4. Return the scale to the upright, normal operating position and make sure it is level. Step 5. Press the ON/OFF key to turn the scale on.2 Step 6.
PD300DHR Digital Height Rod Installation The Digital Height Rod is packed in its own carton inside the main carton. Before starting installation, please unpack carefully and remove all plastic wrappings, foam fillers and cardboard material. You should have the following components: (1) Detecto Digital Height Rod (DHR) (1) Upper Mounting Bracket (with 2 plastic cable clips) (1) Lower Mounting Bracket (1) Hardware Pack, containing: (8) M4x4.5 PM Screws (bracket to height rod) (4) M4x8.
PD300DHR Digital Height Rod Installation, Cont. 3. Align the upper mounting bracket with the holes near the display in the back of the column and install two (2) M4x8.0 PM screws with lock washers to secure the bracket to the column. Align the holes in the lower bracket with the lower holes in the column and install two (2) M4x8.0 PM screws with lock washers to secure it to the column. See Figure No. 21. Figure No. 21 (4) M4x8.0 PM Screws and Lock Washers 4.
PD300MHR Mechanical Height Rod Installation The Mechanical Height Rod is packed in its own carton inside the main carton. Before starting installation, please unpack carefully and remove all plastic wrappings, foam fillers and cardboard material. You should have the following components: (1) Detecto Mechanical Height Rod (MHR) (2) Mounting Brackets (1) Hardware Pack, containing: (4) #8-32x1/2” Pan Head Screws (mounting bar to bracket) (4) #8-32 Hex Nuts (4) M4x8.
PD300MHR Mechanical Height Rod Installation, Cont. 4. Align the upper mounting bracket with the holes near the display in the back of the column and install two (2) M4x8.0 PM screws to secure the bracket to the column. 5. Align the holes in the lower bracket with the lower holes in the column and install two (2) M4x8.0 PM screws with lock washers to secure it to the column. See Figure No. 24. (4) M4x8.0 PM Screws and Lock Washers Figure No. 24 Figure No. 25 6. Referring to Figure No.
BATTERY OPERATION Battery operation is a standard feature of the ProDoc Series Scale, although the batteries are optional (not included). You must first obtain and install six (6) "AA" size Alkaline batteries before operations can begin. Batteries are contained in a battery holder inside the scale. Access is via a removable cover on the bottom of the scale.
BATTERY OPERATION Low Battery When the batteries are near the point they need to be replaced, the display will show BAt. If the battery voltage drops too low for accurate weighing, the scale will automatically shut off and you will be unable to turn it back on. When the BAt message is displayed, the operator should replace the batteries or remove the batteries and plug-in the 9V AC/DC adapter.
OPERATION Keypad Functions Figure No. 31 DO NOT operate the keypad with pointed objects (pencils, pens, etc). Damage to keypad resulting from this practice is NOT covered under warranty. With the scale off, pressing this key will apply power to the scale and turn on the display. If the scale is on, pressing this key will show oFF on the display and turn the scale off. This key is used to reset the display to zero up to the zero limit set during setup and calibration of the scale.
OPERATION Keypad Functions, Cont. The BMI HEIGHT/ENTER key is used to perform several functions during normal operations. Note that the scale will not respond to pressing the key unless the weight is stable. y If the scale has the Digital Height Rod installed, pressing the key once will display the height data from the height rod. Once the height rod is adjusted to the patient, pressing it again will display the Body Mass Index (BMI) calculation.
OPERATION Annunciators The annunciators are turned on to indicate that the scale display is in the mode corresponding to the annunciator label or that the status indicated by the label is active. ZERO This annunciator is turned on to indicate that the weight displayed is within +/- 1/4 division of the center of zero. LOCK This annunciator is turned on to show that the scale is locked onto the displayed weight. STA This annunciator is turned on when the weight on the scale is stable.
OPERATION Basic Weighing 1. Press ON/OFF key. 2. The display will turn on, show all digits and then the software version will be displayed for a few seconds. 3. Next, a small o is shown and moves across the display until a stable zero is established. 4. After establishing zero, the scale will show 0.0 on display. NOTE! If the power up zero is too large, the display will show Err0 for a few seconds, then show off and turn the scale off. 5. Press UNITS key to toggle between kilograms and pounds.
OPERATION Weighing with Height and BMI Measure With Digital Height Rod 1. Press ON/OFF key. 2. The display will turn on, show all digits, the software version for a few seconds and then a small o is shown and moves across the display until a stable zero is established. 3. Next, after establishing zero, the scale will show 0.0 on display. NOTE! If the power up zero is too large, the display will show Err0 for a few seconds, then show off and turn the scale off. 4.
OPERATION Weighing with Height and BMI Measure With Mechanical Height Rod or No Height Rod 1. Press ON/OFF key. 2. The display will turn on, show all digits, the software version for a few seconds and then a small o is shown and moves across the display until a stable zero is established. 3. Next, after establishing zero, the scale will show 0.0 on display. NOTE! If the power up zero is too large, the display will show Err0 for a few seconds, then show off and turn the scale off. 4.
OPERATION Storing Weight with the zSTORE Key 1. With patient on scale and display showing patient’s weight, press zSTORE key. 2. The MEM 1 position annunciator will flash. (Position 1 Selected) 3. Press the zSTORE key until the desired MEM position is selected. The MEM position annunciators (1, 2, 3 or 4) will flash to show which position is currently selected. NOTE! Repeatedly pressing the zSTORE key will advance to the next MEM position and return to the MEM 1 position. (Position 2 Selected) 4.
OPERATION Recalling Weight with the yRECALL Key 1. With the scale in the weight only mode (0.0 on display), press yRECALL key. NOTE! If no weight is stored in a MEM location, then the display will show 0.0. 2. The MEM 1 position annunciator will flash and the stored weight will be displayed. (Position 1 Selected) 3. Press the yRECALL key again. The MEM 2 position annunciator will flash and the stored weight will be displayed. (Position 2 Selected) 4. Press the yRECALL key again.
SETUP Your ProDoc Series Scale has been pre-configured at the factory and should not require changes for use in most applications. However, if the factory settings do not meet the requirements of your operation, the following describes the setup process for your scale. To Enter Setup 1. With the scale off, press and hold the ZERO key and then press the ON/OFF key. 2.
CALIBRATION Your ProDoc Series Scale was calibrated at the factory and should not require adjustment. In the event that the scale should need re-calibration, the following describes the calibration procedure. To maintain the instrument’s high degree of accuracy, a qualified technician should perform this function. Before beginning calibration, the following equipment is required: 150 kg (330 lbs) of calibrated test weight To Enter Calibration 1.
CALIBRATION IMPORTANT! This scale is equipped with an acceleration of gravity constant function which allows the scale to be calibrated in one location and then adjusted to match the acceleration of gravity at the location where it will used. SEtGC y Set Calibrated Gravity Constant The display will alternate between showing the setGC prompt and the current calibrated gravity constant value setting. This setting (e.g. 9.
EVENT COUNTER The ProDoc Series Scale has been designed with an Event Counter type of security seal. When selected, the ProDoc display will show a 3-digit number representing the Calibration and Setup counters. To Review the Event Counter 1. If scale is on, press ON/OFF key. 2. Display will show oFF and turn scale off. 3. Press and hold the UNITS key and then press the ON/OFF key. 4. The display will turn on all digits and show the software version for a few seconds. 5. Release all keys. 6.
DISPLAY MESSAGES The ProDoc Series Scale is equipped with software that tests various portions of the scale's circuitry and verifies proper operation. Should a problem be detected, a message will be displayed alerting the operator to that condition. The following lists the messages displayed and their meaning: Display Meaning Err0 This will be displayed for a few seconds when the scale cannot establish zero and then the display will show off and turn the scale off.
0044-M340-O1 y ProDoc Series Page 32