Cardinal ® Neonatal Scale 6732, 6733, 6735, and 6736 Operation Manual 8525-M039-O1 Rev E 10/03 Cardinal Scale Mfg. Co. P.O. Box 151 Webb City, MO 64870 Ph: 417-673-4631 / Fax: 417-673-5001 web site - www.
TABLE OF CONTENTS Introduction .............................. 1 Features And Specifications ..... 1 Components And Controls ....... 2 Installation ................................ 3 Power Supply ....................... 4 Start-up ................................ 4 Keyboard Functions ................. 4 Operation ................................. 5 Operating Instructions .............. 6 Care And Maintenance ............. 6 Error And Fail Messages .......... 7 Spare Parts Identification .........
INTRODUCTION Thank you for purchasing Detecto Model 6700 Series Neonatal Scale. This instrument has been designed and manufactured in the USA. The Model 6700 Series Neonatal Scale is a solid-state digital scale designed for weighing newborn infants. The Neonatal Scale features a lb/kg key for metric conversion of the infant’s weight. The TARE/ZERO key function subtracts blanket or cover weight on the scale platform allowing the display of infant net weight only.
COMPONENTS AND CONTROLS 2
INSTALLATION 1. Unpack scale and infant tray and note any discrepancies in its condition. Keep carton and packing for return shipment of instrument if necessary. 2. Place scale on a stable, vibration-free location away from direct sunlight and away from any rapidly moving air source. Make certain power cord is stored out of the way of normal traffic. Do not place scale on any unstable cart, stand, or table.
POWER SUPPLY 1. Plug power cord into a fully grounded, 3-pronged wall receptacle that supplies 115 VAC power. If an extension cord must be used, it should be a 3wire, fully grounded type. 2. If a 2-pronged wall receptacle is encountered, it is the responsibility of the customer to contact a qualified electrician to replace the 2-pronged receptacle with a properly grounded 3-pronged wall receptacle or have a grounding adapter properly installed in the 2-pronged receptacle.
OPERATION 5
OPERATING INSTRUCTIONS 1. Place blanket or covering on tray. 2. Press TARE/ZERO key. 3. Place infant on scale. 4. Read weight when display stops flashing. If infant movement causes display to continuously flash, press CLEAR WT. key to lock weight on display. 5. Remove infant and blanket covering from tray. 6. Press CLEAR WT. key. 7. Press TARE/ZERO key. 8. Scale is ready for next use. CARE AND MAINTENANCE 1. DO NOT subject the infant tray to sudden shocks. 2.
ERROR AND FAIL MESSAGES FAIL A This message indicates a bad electrical component requiring technical service. UnStb This message occurs when the infant platform is persistently unstable upon turning on scale. LOD This condition occurs when turning on scale without the infant tray on the scale. UnLOD This message appears when turning on scale with some weight on the infant tray. Remove weight -OL- Indicates scale capacity has been exceeded. -Lo- Indicates gross weight is less than 0.
SPARE PARTS IDENTIFICATION 3 2 4 8 5 1 7 9 6 Item No. 1 Part No. 2950-C014-4A Description LOAD CELL & CONNECTOR ASSEMBLY (SP-50L) 2 427XR1346 BABY TRAY 3 427R1348 RULER FOR BABY SCALE 4 442R137 KEYPAD: 6735 BABY SCALE 5 6540-1011 LEVELER SCREW, 1/4-20x1" LONG, S.S.