Instruction manual
8526-M256-O1 Rev B • CS Series 
15
OPERATIONAL SETUP, CONT. 
0` – Zero Tracking Range 
1.  With the WEIGHT display showing 0` with the 6 flashing, press the 7 key on 
the alphanumeric keypad. 
2.  The WEIGHT display will change to 0` with the 7 flashing. 
3.  Press the MC key. 
4.  The PIECE WEIGHT display will change to show the current setting. 
5.  If the setting displayed is acceptable, press the MC key to save it.  
6.  Otherwise, select the desired setting for the Zero Tracking Range and press the MC 
key to save it. 
 = After weight is stable for over 1 second, it could track ±1/4d (external value is 1/4 bit). 
 = After weight is stable for over 1 second, it could track ±1/2d (external value is 1/2 bit). 
 = After weight is stable for over 1 second, it could track ±1d (external value is 1 bit). 
 = After weight is stable for over 1 second, it could track ±2d (external value is 2 bits). 
NOTE! The zero tracking function will be started only if the Gross weight is at 0. 
0` – Accumulation Ending Mode 
1.  With the WEIGHT display showing 0` with the 7 flashing, press the 8 key on 
the alphanumeric keypad. 
2.  The WEIGHT display will change to 0` with the 8 flashing. 
3.  Press the MC key. 
4.  The PIECE WEIGHT display will change to show the current setting. 
5.  If the setting displayed is acceptable, press the MC key to save it.  
6.  Otherwise, select the desired setting for the Accumulation Ending Mode and press the 
MC key to save it. 
 = Press the M+ key. After screen displays accumulation data for 3 seconds, the scale 
will return to the Weight mode. 
 = Press the M+ key. After screen displays accumulation data, the scale will not
 return 
to the Weight mode until the CE key is pressed. 
 = Press the M+ key. The screen does
 not display accumulation data. The beeper will 
“beep” once. 