Instruction manual
8526-M256-O1 Rev B • CS Series
14
OPERATIONAL SETUP, CONT.
0` – A/D Sampling Speed
1. With the WEIGHT display showing 0` with the 4 flashing, press the 5 key on
the alphanumeric keypad.
2. The WEIGHT display will change to 0` with the 5 flashing.
3. Press the MC key.
4. The PIECE WEIGHT display will change to show the current setting.
5. If the setting displayed is acceptable, press the MC key to save it.
6. Otherwise, select the desired setting for the A/D Sampling Speed and press the MC
key to save it.
– Low Speed
Low speed is about 7.5 Hz. (Weighing reflection is slow but relatively stable).
– Fast Speed
Fast speed is about 14 Hz. (Weighing reflection is fast but relatively not stable).
0` – Zero Display Range
1. With the WEIGHT display showing 0` with the 5 flashing, press the 6 key on
the alphanumeric keypad.
2. The WEIGHT display will change to 0` with the 6 flashing.
3. Press the MC key.
4. The PIECE WEIGHT display will change to show the current setting.
5. If the setting displayed is acceptable, press the MC key to save it.
6. Otherwise, select the desired setting for the Zero Display Range and press the MC
key to save it.
= Display All
= Zero range ± 1 bit will not display division, and displays zero instead.
= Zero range ± 2 bits will not display division, and displays zero instead.
= Zero range ± 3 bits will not display division, and displays zero instead.
If is set, when setting up Pre-Tare data, the Pre-Tare data can not be less or
equal to ±3 bits external value, and so on. When the weight is over 1/3 full capacity
and return to 0, this function will be started.