Specifications
8200-M538-O1 Rev C y 225 Installation & Technical
82
CALIBRATION AND SETUP, CONT.
Fine Span Adjustment
(+0'52#0 (FINE SPAN ADJUSTMENT)
If Fine Span Adjustment is desired, place a calibrated test weight on the scale and with the
SCALE 1 CALIBRATION MENU displayed, press the 4 key and then the ENTER key. The
display will change to show SCALE 1 FINE SPAN ADJUST. Note that #&,:::::::
is the current displayed weight.
5%#.'(+0'52#0#&,756
@72B&190':+6
#&,:::::
y Press UP (Navigation Keys U UP Arrow) to increase the span.
y Press DOWN (Navigation Keys V Down Arrow) to decrease the span.
y Press EXIT (Navigation Keys Y Left Arrow) to save the new setting and return to the
SCALE 1 CALIBRATION MENU.
*+4'5 (HIGH RESOLUTION ADJUSTMENT)
If High Resolution Weight adjustment is desired, place a calibrated test weight on the scale
and with the SCALE 1 CALIBRATION MENU displayed, press the 5 key and then the
ENTER key. The display will change to show SCALE 1 HIRES ADJUSTMENT. Note that
*+4'5:::::: is the active weight in “high resolution” mode (in 1/10 interval).
5%#.'*+4'5#&,756/'06
@72B&190':+6
<'41\GTQUECNG24+06RTKPVYGKIJV
*+4'5::::::
y Press UP (Navigation Keys U UP Arrow) to increase the span.
y Press DOWN (Navigation Keys V Down Arrow) to decrease the span.
y Press EXIT (Navigation Keys Y Left Arrow) to save the new setting and return to the
SCALE 1 CALIBRATION MENU.
y Press the ZERO key to zero the scale.
y Press the PRINT key to print the weight (followed by the text TEST).