8200-M538-O1 Rev C y 225 Installation & Technical
81
CALIBRATION AND SETUP, CONT.
Single-Point for Zero Only (Only Zero) Calibration
%#. (FIRST CALIBRATION WEIGHT)
With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then the
ENTER key. The display will change to show the settings for capacity, interval, decimal
point position and %#.. This is the first of two calibration steps.
1. Insure the scale is empty and then press the ENTER key.
2. The display will flash %#.+$4#6+0) for a few seconds and then change to
show %#..
%#. (SECOND CALIBRATION WEIGHT)
The display will show %#.:::::::. This is the second of two calibration steps.
1. Press the ZERO key.
2. The display will change to show the SCALE 1 CALIBRATION MENU. NOTE! If Span
Adjustment is
required, proceed to the FINE SPAN ADJUSTMENT on page 70.