Specifications

8200-M538-O1 Rev C y 225 Installation & Technical
79
CALIBRATION AND SETUP, CONT.
Dual-Point with Zero (First Zero) Calibration
%#. (FIRST CALIBRATION WEIGHT)
With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then the
ENTER key. The display will change to show the settings for capacity, interval, decimal
point position and %#.. This is the first of two calibration weights. It could be
ZERO (no load) or the TEST load (weight).
1. If the first calibration weight is to be ZERO (no load), press the ENTER key. The display
will flash %#.+$4#6+0) for a few seconds and then change to show %#..
2. If the first calibration weight is to be the TEST load (weight), place the weights on the
scale.
3. Using the numeric keys, enter the value of the TEST load (weight) and then press the
ENTER key.
4. The display will flash %#.+$4#6+0) for a few seconds and then change to
show %#..
%#. (SECOND CALIBRATION WEIGHT)
The display will show %#.:::::::. This is the second of two calibration weights.
It could be ZERO (no load) or the TEST load (weight).
1. If the second calibration weight is to be ZERO (no load), press the 0 key and then the
ENTER key. The display will flash %#.+$4#6+0) for a few seconds and then
change to show %#..
2. If the second calibration weight is to be the TEST load (weight), place the weights on the
scale.
3. Using the numeric keys, enter the value of the test load (weight) and then press the
ENTER key.
4. The display will flash %#.+$4#6+0) for a few seconds and then change to
show %#..
%#. (LAST CALIBRATION WEIGHT)
The display will show %#.. This weight is not used. Press the ENTER key to skip
%#.. The display will change to show the SCALE 1 CALIBRATION MENU. NOTE! If
Span Adjustment is
required, proceed to the FINE SPAN ADJUSTMENT on page 70.