Specifications

8200-M538-O1 Rev C y 225 Installation & Technical
68
CALIBRATION AND SETUP, CONT.
.#$'. – CUSTOM UNITS LABEL
The display will show .#$'.::, where XX is the current setting for the custom units
label. Press the ENTER key to keep the displayed value or use the alphanumeric keys
to enter a new value and press the ENTER key to save the new setting. Note, that only
2 characters are allowed.
<64#%-:: (ZERO TRACKING RANGE)
With SCALE 1 SETUP MENU #1 displayed the current setting for the <64#%-
parameter will be shown. Note that X.X is the current value. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 3 key, the ENTER key and then using the numeric keys, enter the
value for the zero tracking range and then press the ENTER key to save the new setting
and return to SCALE 1 SETUP MENU #1. Allowable values are 0.0 (disables Zero
Tracking), 0.5, 1.0, 2.0, 3.0, 4.0, 5.0, 6.0, 7.0, 8.0 and 9.0.
<.+/+6::: (ZERO LIMIT RANGE)
With SCALE 1 SETUP MENU #1 displayed the current setting for the <.+/+6
parameter will be shown. Note that XXX is the current value. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 4 key, the ENTER key, YES or NO (on the soft keys) and then the
ENTER key to save the new setting and return to the
<.+/+6;'5 <.+/+601
4% (±2%) of scale capacity Full capacity (no limit)
29472<::: (POWER-UP ZERO FEATURE)
With SCALE 1 SETUP MENU #1 displayed the current setting for the 29472<
parameter will be shown. Note that XXX is the current value. If the setting displayed is
acceptable, proceed to the next setup parameter.
Otherwise, press the 4 key, the ENTER key, YES or NO (on the soft keys) and then the
ENTER key to save the new setting and return to the
271;'5 27101
Automatic Re-Zero on Power-Up No Re-Zero on Power-Up