Specifications
8200-M538-O1 Rev C y 225 Installation & Technical
126
TEST MODE AND ERASING MEMORY
Test Mode
The Test mode is used to display the software revision number, the calibration “C” numbers,
and conduct a test of all display elements. The test consists of 6 cycles:
1. With the indicator in normal operations mode, press the SHIFT and then m /ESC key.
2. The display will change to show the (70%6+10 prompt.
3. Press the UNITS key.
4. The display will show the model number and the software revision.
5. Next, a graphics display test will be performed.
6. After the display test, the 225 will display the calibration “C” numbers (C1 to C4) for
approximately 5 seconds.
7. After displaying the “C” numbers, the display will return to normal operation.
Erase Ram Memory
WARNING! This procedure will ERASE ALL of the internal RAM memory! This
includes the ID file, batcher info, dealer string, mode of operation, visual tickets).
If a 225 does not display properly after a program update or otherwise becomes corrupt,
perform the following:
1. With the indicator in normal operations mode, press the SHIFT key and then the
Navigation ENTER key (red square key).
2. The display will change to show the SETUP/REVIEW MENU.
3. Using the numeric keys, enter 130 and then press the ENTER key.
4. The display will change to show a warning message.
5. To ERASE ALL the memory, press YES (on the soft keys) and then the ENTER key.
6. Otherwise, press NO (on the soft keys) and then the ENTER key to return the
SETUP/REVIEW MENU.
Erasing EEPROM Memory
WARNING! This procedure will ERASE the EEPROM calibration data. All EEPROM
memory will be erased and the settings for the scales, Serial ports, etc. will be set
to their default values.
1. With the indicator in normal operations mode, press the SHIFT key and then the
Navigation ENTER key (red square key).
2. The display will change to show the SETUP/REVIEW MENU.
3. Using the numeric keys, enter 160 and then press the ENTER key.
4. The display will change to show a warning message.
5. To ERASE the calibration data, press YES (on the soft keys) and then the ENTER key.
6. Otherwise, press NO (on the soft keys) and then the ENTER key to return the
SETUP/REVIEW MENU.