Product Manual

D Electromagnetic Emissions Standards Compliance
Contents
Guidance and manufacturers declaration—electromagnetic emissions ..................................D-1
Guidance and manufacturers declaration—electromagnetic immunity ..................................D-1
Recommended separation distances between portable and mobile RF
communications equipment and the AED .............................................................................................D-3
Guidance and manufacturers declaration—electromagnetic emissions
The AED is intended for use in the electromagnetic environment specied below. The customer or the user of the AED should assure that it is
used in such an environment.
Emissions test Compliance Electromagnetic environmentguidance
RF emissions
CISPR 11
Group 1 The AED uses RF energy only for its internal function. Therefore its RF emissions are very low and
are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B The AED is suitable for use in all establishments, including domestic establishments and those
directly connected to the public low-voltage power supply network that supplies buildings
Harmonic emissions used for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage uctuations/
icker emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturers declaration—electromagnetic immunity
The AED is intended for use in the electromagnetic environment specied below. The customer or the user of the AED should assure that it is
used in such an environment.
Immunity test IEC 60601 test level
Compliance
level
Electromagnetic
environment—guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If oors
are covered with synthetic material, the relative humidity
should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
Not applicable
Surge
IEC 61000-4-5
±1 kV dierential mode
±2 kV common mode
Not applicable
D-1Powerheart® G5 Automated External Debrillator 70-01704-01 B