Product Manual
B Technical Data
Contents
◆ Powerheart G5 parameters ........................................................... B-1
◆ Debrillation pads ............................................................................B-3
◆ Intellisense® battery (model XBTAED001) ................................B-3
This section lists the physical, operating, standby, and storage parameters of the AED, and the physical parameters of the debrillation
pads and AED battery.
Table 2: Environmental information (continued)
Parameter Detail
Pressure • CSA evaluated: 700 hPa to 1060 hPa
• Minimum: 570 hPa
• Maximum: 1060 hPa
* Operating: AED with pads and battery installed and lid open.
** Standby: AED with pads and battery installed and lid closed.
*** Storage and transport: AED with pads optionally connected and battery not installed.
****The debrillator and its accessories have been tested for long-term standby for 24
months in the temperature range from 20 to 30 °C only. For operation, namely, when
the debrillator lid is open, the operating temperature range is from 0 to 50 °C. Hence,
after each rescue attempt, please close the lid, thus bringing the debrillator back to
the standby state, with the associated standby requirement of 20 to 30 °C only. The
debrillator has not been qualied for long term standby in an environment between
30 to 50 °C.
The AED can be stored in long-term standby at temperatures between 0 to 50 °C.
However, continuous storage at temperature extremes may reduce the life of the
electrodes and / or battery. Rescue Ready technology assesses the AED battery and
pads to ensure they are ready to perform a rescue. When the Rescue Ready status is
green, the AED is ready for a rescue. Audible and visual indicators will notify when
the device requires attention. Deployments in the conditions described above are
covered under the Cardiac Science AED warranty, battery performance replacement
guarantee, electrode warranty through expiration date and Cardiac Science indemni-
cation agreement. Refer to Section F for complete warranty information.
Powerheart G5 parameters
Table 1: Physical parameters
Parameter Detail
Operation • Automatic
• Semi-automatic
• Multi-language (in specied combinations only)
Dimensions • Height: 9 cm (3.4 in)
• Width: 23 cm (9.0 in)
• Depth: 30 cm (11.8 in)
Weight (with
battery and pads)
2.6 kg (5.7 lb)
Table 2: Environmental information
Parameter Detail
Operating* • Temperature: 0°C to 50°C (32°F to 122°F)
• Humidity: 10% to 95% (non-condensing)
Standby** • Short-term (5 days) temperature:
0° C to 50° C (32° F to 122° F)
• Long-term temperature****:
20° C to 30° C (68° F to 86° F)
• Humidity: 10% to 95% (non-condensing)
Storage and
transport
(up to 3 days)***
• Temperature: -30°C to 65°C (-22°F to 149°F)
• Humidity: 10% to 95% (non-condensing)
Altitude • CSA evaluated: -382 m to 3000 m
• Minimum: -382 m
(approximate; calculated from pressure)
• Maximum: 4594 m
(approximate; calculated from pressure)
B-1Powerheart® G5 Automated External Debrillator 70-01704-01 B